Endoscopes – What is the problem?

Glenys Harrington

 

Gastrointestinal endoscopes are complex devices that can be difficult to clean and disinfect. Over the last few years there have been increasing reports of outbreaks associated with endoscopes in particular duodenoscopes which are used for endoscopic retrograde cholangiopancreatography (ERCP) procedures.

These outbreaks have been associated with multidrug resistant organism’s including Carbapenem-resistant Enterobacteriaceae (CRE). Some of these outbreaks occurred in settings where the manufacturers’ instructions had been followed correctly and others in settings of inadequate cleaning, inappropriate disinfection and design flaws associated with the endoscopes.

Gastrointestinal endoscopes become heavily contaminated (107 – 1010 bacteria) during use. Researchers have shown that the cleaning step results in a 2-6 log10 reduction of microbes and the high level disinfection step results in another 4-6 log10 reduction.  Hence the margin of safety is minimal and if there is any deviation from correct reprocessing procedures it may result in the endoscope remaining contaminated and possible patient-to-patient transmission.

About the College

The ACIPC is the peak body for Infection Prevention and Control professionals in the Australasian region. Our stated vision is the prevention and control of infection in our communities. We commenced in January 2012 bringing together the various State and Territory infection control associations formerly in AICA (The Australian Infection Control Association) to support and encourage collaboration across Australasia.

Conference Managers

Please contact the team at Conference Design with any questions regarding the conference.
© 2015 - 2016 Conference Design Pty Ltd